A medical device manufacturer complained of intermittent memory failures that caused their medical device to hang or crash. The failures were troublesome but random and not easy to isolate. The team looked for months but could not isolate the failures enough to find root cause. FuturePlus Systems to the rescue! Within 3 days we found the problem. The BIOS was not setting ODT correctly in the DRAMs Mode Registers. It turns out the BIOS was recently updated and a bug was introduced that was not setting up the DIMMs correctly.
Here is a picture of what a few of the DDR3 Memory data signals looked like.
Vive la difference!!! So what’s in your medical device system? Don’t be liable for not doing ‘due diligence’ on every part of your product. FuturePlus Systems offers a Memory Subsystem Audit to ensure quality and proper operation. Our services are far less expensive than buying the equipment and doing it yourself. Plus you are getting experts in Memory Validation with a 25+ year track record for excellence.